14
Command
Return
SIwxyz
none
Function:
This command sets the integration time for acquiring a count from the voltage-to-
frequency converter. The four-digit integer value of wxyz is equal to the total
integration time in units of 0.01 seconds. Thus, a value of 0100 will set the
integration time to 1.00 seconds.
Warning:
Values less than 0100 (1.00 seconds) are of limited value because the rise time and
noise of the system electronics demand longer integrations. The maximum count
that can be retained is 2 bytes wide (2
16
), which will limit long integration times.
Table 4-3. Set Integration Time Command.
Command
Return
SCOUNT
C=vwxyz CR LF
Function:
This command starts an integration for the selected integration time and gain
settings. The vwxyz output value is the integer count. The left side of the value is
padded with zeros so that a five-character value is always sent. Thus, a count of
100 is sent as C=00100 CR LF.
Warning:
The maximum value that can be sent is two bytes wide or 65536 decimal. During
the integration time, the SSP-4 will not accept any serial inputs.
Table 4-4. Start Count Command