Imaging Modalities
2D/BMode Imaging
7-44 Revision D.0
Following are some guidelines for using SONOS biopsy controls:
If the selected probe supports multiple needle mounting positions, you can
use a rotary control to select crossover depth. (See page 7-42.)
Biopsy graphics are erased from the screen in cardiac presets, and in PW,
CW, MMode, and zoom modes.
Changing the probe selection turns off biopsy support; it can be reactivated
only when a probe that accepts biopsy guides is selected.
When a biopsy screen display is on, the trackball controls a marker
indicating the position on the image that is the target of the biopsy procedure
along the needle path. The minimum needle length required to reach the
target marker is displayed in the upper right corner of the screen.
Probes supporting biopsy
The following probes accept biopsy needle guides.
To order biopsy kits contact
CIVCO Medical Instruments Co.
102 First Street
Kalona, IA 52247-9589 U.S.A.
www.civco.com
Phone inside US: (800) 445-6741
Phone outside US: 1-319-656-4447
Fax orders to: 1-319-656-4451
For information about ordering other Philips Medical Systems supplies and
accessories, see “Supplies and Accessories” on page 1-17.
NOTE See the Transducer Reference for safety and maintenance information about the
use of biopsy needles and guides.
Probe Model CIVCO Biopsy Kit Crossover Depths
E6509 21336A 610-588 187 mm
c3540 21321A 645-057 90 mm
C5040 21373A 645-032 20 mm
50 mm