Imaging Modalities
Angio Imaging
7-118 Revision D.0
Angio Troubleshooting
Symptoms Suggestions
Not sensitive. Scale.
Adjust Angio Gain; it could be too low.
Adjust Focus to point at the area of interest.
Choose a filter to reduce flash or improve sensitivity.
Reposition the transducers, or try another imaging
window.
Power.
Turn Angio Priority on for small vessels.
Adjust Smoothing.
Need a higher
frame rate.
Decrease the width of the Angio area by pressing
and using the trackball.
For sector or curvilinear array transducers, remove the
grayscale image outside the color area by turning on
B/W Suppress.
Density
Packet size
Scale.
Size